Título

Analytical assessment of sodium ISFET based sensors for sweat analysis

AutorRovira, Meritxell; Lafaye, Celine; Wang, Shu; Fernández Sánchez, César CSIC ORCID ; Saubade, Mathieu; Liu, Shih Chii; Jiménez-Jorquera, Cecilia CSIC ORCID CVN
Palabras claveSensor characterization | Sodium ISFETs | Sweat analysis
Fecha de publicación15-oct-2023
EditorElsevier
CitaciónSensors and Actuators B: Chemical
ResumenBody thermoregulation during exercise induces sweating with the consequent loss of water, electrolytes and other compounds. The use of sweat for bioanalytical purposes has recently widespread because it is an easily accessible biofluid that can be noninvasively collected and/or directly monitored using wearable devices. Different sweat biomarkers can be monitored as indicators of the physiological status of an individual. The concentration of sweat Na+ electrolyte provides information about dehydration or hyponatremia events. Among the different sensor technologies applied for Na+ ion detection in sweat, ion-sensitive field-effect transistors (ISFETs) show superior features in terms miniaturization, key for working with the low volumes of sweat available, robustness, scalability and reproducibility. They also offer fast response and low impedance output signal. This work reports an in-depth study that thoroughly assesses the potential of ISFET sensors for sweat Na+ analysis. Results show a reproducible sensitivity of 60.7 ± 0.5 mV (-Log aNa)− 1, high repeatability, and lifetime up to one month. Sensor reliability is demonstrated by analysing 20 sweat samples and results are compared with the ones provided with the standard ion chromatography technique (IC). The statistical analysis demonstrates that Na+ concentrations estimated with the ISFET sensor and IC are in good agreement showing a relative error of up to 20%. Results demonstrate that the sensor presented in this work can potentially be used for the continuous monitoring of Na+ changes in sweat during exercise.
Versión del editorhttps://doi.org/10.1016/j.snb.2023.134135
URIhttp://hdl.handle.net/10261/351436
DOI10.1016/j.snb.2023.134135
ISSN09254005
Aparece en las colecciones: (IMB-CNM) Artículos

Ficheros en este ítem:
Fichero Descripción Tamaño Formato
1-s2.0-S092540052300850X-main.pdfArtículo principal3,05 MBAdobe PDFVisualizar/Abrir
Mostrar el registro completo

CORE Recommender

SCOPUSTM   
Citations

2
checked on 14-may-2024

Page view(s)

19
checked on 22-may-2024

Download(s)

3
checked on 22-may-2024

Google ScholarTM

Check

Altmetric

Altmetric


Este item está licenciado bajo una Licencia Creative Commons Creative Commons