English   español  
Please use this identifier to cite or link to this item: http://hdl.handle.net/10261/219084
Share/Impact:
Statistics
logo share SHARE logo core CORE   Add this article to your Mendeley library MendeleyBASE

Visualizar otros formatos: MARC | Dublin Core | RDF | ORE | MODS | METS | DIDL
Exportar a otros formatos:

Title

A new lower limb portable exoskeleton for gait assistance in neurological patients: a proof of concept study

AuthorsPuyuelo-Quintana, G.; Cano-de-la-Cuerda, R.; Plaza-Flores, A.; Garces-Castellote, E.; Sanz-Merodio, D.; Goñi-Arana, A.; Marín-Ojea, J.; García-Armada, E.
KeywordsExoskeletons
Feasibility
Gait
Multiple sclerosis
Safety
Stroke
Issue Date6-May-2020
PublisherBioMed Central
CitationJournal of NeuroEngineering and Rehabilitation 17 (1): 60 (2020)
AbstractBackground: Few portable exoskeletons following the assist-as-needed concept have been developed for patients with neurological disorders. Thus, the main objectives of this proof-of-concept study were 1) to explore the safety and feasibility of an exoskeleton for gait rehabilitation in stroke and multiple sclerosis patients, 2) to test different algorithms for gait assistance and measure the resulting gait changes and 3) to evaluate the user's perception of the device. Methods: A cross-sectional study was conducted. Five patients were recruited (4 patients with stroke and 1 with multiple sclerosis). A robotic, one-degree-of-freedom, portable lower limb exoskeleton known as the Marsi Active Knee (MAK) was designed. Three control modes (the Zero Force Control mode, Mode 1 and Mode 3) were implemented. Spatiotemporal gait parameters were measured by the 10-m walking test (10MWT), the Gait Assessment and Intervention Tool (G.A.I.T.) and Tinetti Performance Oriented Mobility Assessment (gait subscale) before and after the trials. A modified QUEST 2.0 questionnaire was administered to determine each participant's opinion about the exoskeleton. The data acquired by the MAK sensors were normalized to a gait cycle, and adverse effects were recorded. Results: The MAK exoskeleton was used successfully without any adverse effects. Better outcomes were obtained in the 10MWT and G.A.I.T. when Mode 3 was applied compared with not wearing the device at all. In 2 participants, Mode 3 worsened the results. Additionally, Mode 3 seemed to improve the 10MWT and G.A.I.T. outcomes to a greater extent than Mode 1. The overall score for the user perception of the device was 2.8 ± 0.4 95% CI. Conclusions: The MAK exoskeleton seems to afford positive preliminary results regarding safety, feasibility, and user acceptance. The efficacy of the MAK should be studied in future studies, and more advanced improvements in safety must be implemented.
Publisher version (URL)https://doi.org/10.1186/s12984-020-00690-6
URIhttp://hdl.handle.net/10261/219084
DOI10.1186/s12984-020-00690-6
E-ISSN1743-0003
Appears in Collections:(CAR) Artículos
Files in This Item:
File Description SizeFormat 
A_new_lower_limb_portable_exoskeleton.pdfArtículo principal2,42 MBAdobe PDFThumbnail