Por favor, use este identificador para citar o enlazar a este item: http://hdl.handle.net/10261/193942
COMPARTIR / EXPORTAR:
logo share SHARE logo core CORE BASE
Visualizar otros formatos: MARC | Dublin Core | RDF | ORE | MODS | METS | DIDL | DATACITE

Invitar a revisión por pares abierta
Título

Experimental validations of a tunable-lens-based visual demonstrator of multifocal corrections

AutorAkondi, Vyas CSIC ORCID; Sawides, L. CSIC ORCID; Marrakchi, Y.; Gambra, Enrique CSIC ORCID; Marcos, Susana CSIC ORCID ; Dorronsoro, Carlos CSIC ORCID
Fecha de publicación15-nov-2018
EditorOptical Society of America
CitaciónBiomedical Optics Express 9(12): 6302-6317 (2018)
ResumenThe Simultaneous Vision simulator (SimVis) is a visual demonstrator of multifocal lens designs for prospective intraocular lens replacement surgery patients and contact lens wearers. This programmable device employs a fast tunable lens and works on the principle of temporal multiplexing. The SimVis input signal is tailored to mimic the optical quality of the multifocal lens using the theoretical SimVis temporal profile, which is evaluated from the through-focus Visual Strehl ratio metric of the multifocal lens. In this paper, for the first time, focimeter-verified on-bench validations of multifocal simulations using SimVis are presented. Two steps are identified as being critical to accurate SimVis simulations. Firstly, a new iterative approach is presented that improves the accuracy of the theoretical SimVis temporal profile for three different multifocal intraocular lens designs – diffractive trifocal, refractive segmented bifocal, and refractive extended depth of focus, while retaining a low sampling. Secondly, a fast focimeter is used to measure the step response of the tunable lens, and the input signal is corrected to include the effects of the transient behavior of the tunable lens. It was found that the root-mean-square of the difference between the estimated through-focus Visual Strehl ratio of the multifocal lens and SimVis is not greater than 0.02 for all the tested multifocal designs.
Descripción16 pags., 8 figs. 2 tabs.
Versión del editorhttps://doi.org/10.1364/BOE.9.006302
URIhttp://hdl.handle.net/10261/193942
DOI10.1364/BOE.9.006302
Identificadoresdoi: 10.1364/BOE.9.006302
issn: 2156-7085
Aparece en las colecciones: (CFMAC-IO) Artículos




Ficheros en este ítem:
Fichero Descripción Tamaño Formato
Experimental validations.pdf3,65 MBAdobe PDFVista previa
Visualizar/Abrir
Mostrar el registro completo

CORE Recommender

PubMed Central
Citations

6
checked on 14-abr-2024

SCOPUSTM   
Citations

14
checked on 16-abr-2024

WEB OF SCIENCETM
Citations

14
checked on 29-feb-2024

Page view(s)

208
checked on 19-abr-2024

Download(s)

189
checked on 19-abr-2024

Google ScholarTM

Check

Altmetric

Altmetric


Artículos relacionados:


NOTA: Los ítems de Digital.CSIC están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.