Cryocooler-based helium recovery plant for applications requiring gas or liquid with extreme purity

AutorGabal Lanau, Miguel; Rillo, Conrado; Sesé Monclús, Javier ; Diederichs, J.; Rayner, G.; Spagna, S.
Fecha de publicación2016
CitaciónICC 19 (2016)
ResumenA cryocooler based helium recovery plant designed to store, compress and purify helium gas collected from one or more helium-using instruments, as well as to liquefy and redistribute the purified gas within a closed system, is presented. The recovery plant is composed to match the purification and liquefaction rate of the system with the consumption rate of the coupled instruments in an attempt to achieve zero loss. In addition, extreme purity helium (gas and liquid) is obtained by the combination of cryo-condensation of impurities and further chemical adsorption by a getter. Problems like clogging of low temperature impedances by impurities normally present in helium have been definitely solved using this plant concept.
DescripciónTrabajo presentado al 19th International Cryocooler Conference, celebrada en San Diego, California (US) del 20 al 23 de junio de 2016.
URIhttp://hdl.handle.net/10261/148801
Aparece en las colecciones: (ICMA) Comunicaciones congresos
Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
cryocoolerpurity.pdf659,38 kBAdobe PDFVista previa
Visualizar/Abrir
Mostrar el registro completo
 


NOTA: Los ítems de Digital.CSIC están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.