Frequency response of an atomic force microscope in liquids and air: Magnetic versus acoustic excitation

AutorHerruzo, Elena T. ; García García, Ricardo
Palabras claveAmplitude modulation
Atomic force microscopy
Cantilevers
Micromechanical devices
Fecha de publicación2-oct-2007
EditorAmerican Institute of Physics
CitaciónApplied Physics Letters 91, 143113 (2007)
ResumenWe discuss the dynamics of an amplitude modulation atomic force microscope in different environments such as water and air. Experiments, analytical expressions, and numerical simulations show that the resonance curves depend on the excitation method used to drive the cantilever, either mechanical or magnetic. This dependence is magnified for small force constants and quality factors, i.e., below 1 N/m and 10, respectively. We show that the equation for the observable, the cantilever deflection, depends on the excitation method. Under mechanical excitation, the deflection involves the base and tip displacements, while in magnetic excitation, the cantilever deflection and tip displacement coincide
Versión del editorhttp://link.aip.org
http://dx.doi.org/10.1063/1.2794426
URIhttp://hdl.handle.net/10261/13781
DOI10.1063/1.2794426
ISSN0003-6951
Aparece en las colecciones: (IMN-CNM) Artículos
Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
Herruzo, Elena T. and García, Ricardo Appl.Phys.Lett._91_2007.pdf202,53 kBAdobe PDFVista previa
Visualizar/Abrir
Mostrar el registro completo
 

Artículos relacionados:


NOTA: Los ítems de Digital.CSIC están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.