English   español  
Por favor, use este identificador para citar o enlazar a este item: http://hdl.handle.net/10261/11866
Compartir / Impacto:
Estadísticas
Add this article to your Mendeley library MendeleyBASE
Citado 34 veces en Web of Knowledge®  |  Pub MebCentral Ver citas en PubMed Central  |  Ver citas en Google académico
Visualizar otros formatos: MARC | Dublin Core | RDF | ORE | MODS | METS | DIDL
Título

Fast and simultaneous monitoring of organic pollutants in a drinking water treatment plant by a multi-analyte biosensor followed by LC–MS validation

AutorRodríguez-Mozaz, Sara; López de Alda, Miren; Barceló, Damià
Palabras claveImmunosensor
TIRF
Multi-analyte
Pesticides
Estrogens
Water
Waterworks
Fecha de publicación15-abr-2006
EditorElsevier
CitaciónTalanta 69(2):377-384 (2006)
ResumenThis work describes the application of an optical biosensor (RIver ANALyser, RIANA) to the simultaneous analysis of three relevant environmental organic pollutants, namely, the pesticides atrazine and isoproturon and the estrogen estrone, in real water samples. This biosensor is based on an indirect inhibition immunoassay which takes place at a chemically modified optical transducer chip. The spatially resolved modification of the transducer surface allows the simultaneous determination of selected target analytes by means of “total internal reflection fluorescence” (TIRF). The performance of the immunosensor method developed was evaluated against a well accepted traditional method based on solid-phase extraction followed by liquid chromatography–mass spectrometry (LC–MS). The chromatographic method was superior in terms of linearity, sensitivity and accuracy, and the biosensor method in terms of repeatability, speed, cost and automation. The application of both methods in parallel to determine the occurrence and removal of atrazine, isoproturon and estrone throughout the treatment process (sand filtration, ozonation, activated carbon filtration and chlorination) in a waterworks showed an overestimation of results in the case of the biosensor, which was partially attributed to matrix and cross-reactivity effects, in spite of the addition of ovalbumin to the sample to minimize matrix interferences. Based on the comparative performance of both techniques, the biosensor emerges as a suitable tool for fast, simple and automated screening of water pollutants without sample pretreatment. To the author's knowledge, this is the first description of the application of the biosensor RIANA in the multi-analyte configuration to the regular monitoring of pollutants in a waterworks
Descripción8 pages, 2 tables, 3 figures.
Versión del editorhttp://dx.doi.org/10.1016/j.talanta.2005.09.050
URIhttp://hdl.handle.net/10261/11866
DOI10.1016/j.talanta.2005.09.050
ISSN0039-9140
Aparece en las colecciones: (IDAEA) Artículos
Ficheros en este ítem:
No hay ficheros asociados a este ítem.
Mostrar el registro completo
 



NOTA: Los ítems de Digital.CSIC están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.