Part per trillion determination of atrazine in natural water samples by a surface plasmon resonance immunosensor

AutorFarré, Marinella CSIC ORCID; Martínez, Elena CSIC ORCID ; Ramón-Azcón, Javier CSIC ORCID; Navarro-Ortega, Alicia; Radjenovic, Jelena CSIC ORCID; Mauriz, Elba CSIC ORCID; Lechuga, Laura M. CSIC ORCID ; Marco, María Pilar CSIC ORCID CVN ; Barceló, Damià CSIC ORCID
Palabras claveBiosensors
SPR inmunosensor
Atrazine
Fecha de publicación13-mar-2007
EditorSpringer Nature
CitaciónAnalytical and Bioanalytical Chemistry 388(1): 207-214 (2007)
ResumenA new immunoassay for continuously monitoring atrazine in water has been developed. It uses a portable biosensor platform based on surface plasmon resonance (SPR) technology. This immunoassay is based on the binding inhibition format with purified polyclonal antibodies, with the analyte derivative covalently immobilized on a gold sensor surface. An alkanethiol self-assembled monolayer (SAM) was formed on the gold-coated sensor surface in order to obtain a reusable sensing surface. The low detection limit for the optimized assay, calculated as the concentration that produces a 10% decrease in the blank signal, is 20 ng/L. A complete assay cycle, including regeneration, is accomplished in 25 min. Additionally, a study of the matrix effects of different types of wastewater was performed. All measurements were carried out with the SPR sensor system (β-SPR) commercialised by the company Sensia, S.L. (Spain). The small size and low response time of the β-SPR platform would allow it to be used in real contaminated locations. The immunosensor was evaluated and validated by measuring the atrazine content of 26 natural samples collected from Ebro River. Solid-phase extraction followed by gas chromatography coupled to mass spectrometric detection (SPE–GC–MS) was used to validate the new immunoassay.
Descripción8 pages, 5 figures.-- PMID: 17393154 [PubMed].-- Printed version published in May 2007.
Versión del editorhttp://dx.doi.org/10.1007/s00216-007-1214-2
URIhttp://hdl.handle.net/10261/9447
DOI10.1007/s00216-007-1214-2
ISSN1618-2642
E-ISSN1618-2650
Aparece en las colecciones: (IQAC) Artículos
(IMN-CNM) Artículos

Mostrar el registro completo

CORE Recommender

SCOPUSTM   
Citations

95
checked on 04-abr-2024

WEB OF SCIENCETM
Citations

84
checked on 23-feb-2024

Page view(s)

358
checked on 19-abr-2024

Google ScholarTM

Check

Altmetric

Altmetric


NOTA: Los ítems de Digital.CSIC están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.