Please use this identifier to cite or link to this item: http://hdl.handle.net/10261/27874
Share/Export:
logo share SHARE logo core CORE BASE
Visualizar otros formatos: MARC | Dublin Core | RDF | ORE | MODS | METS | DIDL | DATACITE

Invite to open peer review
Title

Nanomechanics for specific biological detection

AuthorsÁlvarez, Mar CSIC ORCID; Carrascosa, Laura G. CSIC ORCID; Tamayo de Miguel, Francisco Javier CSIC ORCID; Calle Martín, Ana CSIC; Lechuga, Laura M. CSIC ORCID
KeywordsNanotechnology
Biosensors
Microcantilevers
Nanomechanics
DNA
Biochips
Issue Date28-Jul-2003
PublisherThe International Society for Optics and Photonics
CitationProceedings of SPIE 5118: 197-206 (2003)
AbstractNanomechanical biosensors have emerged as a promising platform for specific biological. Among the advantages are direct detection without need of labelling with fluorescent or radioactive molecules, very high sensitivity, reduced sensor area, and suitability for integration using silicon technology. Here we have studied the immobilization of oligonucleotide monolayers by monitoring the microcantilever bending. Oligonucleotides were derivatized with thiol molecules for self-assembly on the gold-coated side of a microcantilever. The geometry of the binding and the surface density were studied by mixing derivatized oligonucleotides with spacer self-assembled monolayers and by controlling the oligonucleotide functional group form. These results are compared with fluoresencent and chemiluminescence techniques. Furthermore, we present the first results of direct pesticide detection with microcantilever-based biosensors. Herbicide DDT was detected by performing competitive assays, in which the cantilever was coated with a synthetic DDT hapten, and it was exposured to different rations between the monoclonal antibody and the DDT. A new technique is presented for the detection of the nanomechanical response for biosensing applications, in which the resonant frequency is measured with about two orders of magnitude higher sensitivity. The low quality factor of the microcantilever in liquid is increased up by using an active feedback control, in which the cantilever oscillation is amplified and delayed and it is used as a driving force. The technique has been applied for the detection of ethanol, proteins, and pathogens.
Description10 páginas, 6 figuras.-- Trabajo presentado en la conferencia "Nanotechnology"; Maspalomas, Gran Canaria (España); 19-Mayo-2003; Editores: Robert Vajtai, Xavier Aymerich, Laszlo B. Kish, Angel Rubio.
Publisher version (URL)http://dx.doi.org/10.1117/12.498629
URIhttp://hdl.handle.net/10261/27874
DOI10.1117/12.498629
ISSN0277-786X
Appears in Collections:(IMN-CNM) Artículos




Files in This Item:
File Description SizeFormat
GetPDFServlet.pdf244,12 kBAdobe PDFThumbnail
View/Open
Show full item record

CORE Recommender

SCOPUSTM   
Citations

5
checked on Apr 13, 2024

WEB OF SCIENCETM
Citations

5
checked on Feb 22, 2024

Page view(s)

433
checked on Apr 19, 2024

Download(s)

222
checked on Apr 19, 2024

Google ScholarTM

Check

Altmetric

Altmetric


WARNING: Items in Digital.CSIC are protected by copyright, with all rights reserved, unless otherwise indicated.