Título : M-H loop tracer based on digital signal processing for low frequency characterization of extremely thin magnetic wires
Autor : Butta, Mattia, Infante, Germán, Ripka, Pavel, Badini-Confalonieri, G. A., Vázquez, M.
Fecha de publicación : 21-Aug-2009
Editor: American Institute of Physics
Citación : Review of Scientific Instruments 80(8): 083906 (2009)
Resumen: A high-sensitivity ac hysteresis loop tracer has been developed to measure the low frequency hysteresis loop of soft magnetic materials. It has been applied successfully to characterize straight pieces of amorphous glass-covered microwires with metallic nucleus down to 1.5 µm thick. Based on the electromagnetic induction law, the proposed design is extremely simple and exploits the capabilities of commercially available data acquisition cards together with digital signal processing in order to achieve high-sensitivity without the need of expensive analog equipment.
Descripción : 4 pages, 4 figures.
Versión del editor: http://dx.doi.org/10.1063/1.3206264
URI : http://hdl.handle.net/10261/21355
ISSN: 0034-6748