Título: | Direct and simultaneous determination of reduced and oxidized glutathione and homoglutathione by liquid chromatography-electrospray/mass spectrometry in plant tissue extracts |
Autor: | Rellán-Álvarez, Rubén CSIC ORCID; Hernández, Luis E.; Abadía Bayona, Javier CSIC ORCID ; Álvarez-Fernández, Ana CSIC ORCID | Palabras clave: | Ascorbate Glutathione Homoglutathione Liquid chromatography Mass spectrometry Oxidized glutathione S-Nitrosoglutathione |
Fecha de publicación: | 12-jun-2006 | Editor: | Elsevier | Citación: | Analytical Biochemistry 356(2): 254-264 (2006) | Resumen: | A simple, highly selective, sensitive, and reproducible liquid chromatography–electrospray ionization/mass spectrometry (time of flight) method has been developed for the direct and simultaneous determination of glutathione and related compounds such as homoglutathione in different plant tissues. These compounds are low-molecular mass antioxidants involved in cellular redox homeostasis in plants, and efforts are being made to develop methods to determine the concentrations of oxidized and reduced forms of these compounds and their ratio. Many of the methodologies developed so far, however, are time-consuming and complex; therefore, analytes can decompose and their redox status can change during the analysis process. The method we have developed allows the simultaneous determination of reduced forms (glutathione [GSH] and homoglutathione [hGSH]) and oxidized forms (glutathione disulfide [GSSG]) of these compounds and is also suitable for the determination of ascorbic acid (ASA) and S-nitrosoglutathione (GSNO). Quantification was done using isotopically labeled GSH and ASA as internal standards. All compounds were base peak resolved in less than 6 min, and limits of detection were 60 pmol for GSH, 30 pmol for hGSH, 20 pmol for GSSG, 100 pmol for ASA, and 30 pmol for GSNO. The intraday repeatability values were approximately 0.4 and 7% for retention time and peak area, respectively, whereas the interday repeatability values were approximately 0.6 and 9% for retention time and peak area, respectively. Analyte recoveries found were between 92 and 105%. The method was used to determine the concentrations of GSH, GSSG, hGSH, and ASA in extracts from several plant tissues. | Descripción: | 11 pages, 5 figures, 4 tables.-- PMID: 16828049 [PubMed].-- Printed version published Sep 15, 2006. | Versión del editor: | http://dx.doi.org/10.1016/j.ab.2006.05.032 | URI: | http://hdl.handle.net/10261/17268 | DOI: | 10.1016/j.ab.2006.05.032 | ISSN: | 0003-2697 |
Aparece en las colecciones: | (EEAD) Artículos |
Mostrar el registro completo
CORE Recommender
SCOPUSTM
Citations
87
checked on 18-abr-2024
WEB OF SCIENCETM
Citations
83
checked on 26-feb-2024
Page view(s)
361
checked on 24-abr-2024
Google ScholarTM
Check
Altmetric
Altmetric
NOTA: Los ítems de Digital.CSIC están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.